Поставляем Датчик расстояния OD Precision OD Precision - OD5-85W20 - 6035980 для промышленных и гражданских объектов. Оригинальная продукция с гарантией. Подбор аналога, доставка по Москве, СПб и регионам РФ.
| System part: | Sensor head |
|---|
| Measuring range: | 65 mm ... 105 mm, 6 % ... 90 % remission; at default settings |
|---|---|
| Resolution 1): | 1 µm |
| Repeatability 2): | 3 µm |
| Linearity 3): | ± 20 µm |
| Response time 4) 5): | 0.1 ms |
| Measuring frequency: | 10 kHz 6) |
| Light source: | Laser, red |
| Typ. light spot size (distance): | 260 µm x 1,200 µm (85 mm) |
| Additional function: | Automatic sensitivity adjustment, Selectable measuring frequency (automatic / 0.1 ms ... 3.2 ms), Anti interference mode, Glass thickness measurement, Manual sensitivity adjustment, Averaging settings 1 ... 4,096x |
| Note: | OD Precision sensor head can be used with AOD5-P/N1 or stand-alone via RS-422 |
| Laser class: | 2 (EN 60825-1) 7) |
| Output time: | ? 0.1 ms |
| footnote | 1) 3) Measurement on 90 % remission (ceramic, white), for OD5-25x measurement on mirror; averaging set to: 4096 2) Measurement on 90 % remission (ceramic, white), for OD5-25x measurement on mirror; averaging set to: 4096; constant ambient conditions 4) Time needed for automatic sensitivity adjustment is calculated as: sampling period x 20. At default setting 100 µs (10kHz) this is |
| Data interface: | RS-422 |
|---|---|
| Laser off input: | 1 x laser-off |
| Electrical connection 1): | 0.5 m cable with connector |
|---|---|
| Supply voltage Vs 2): | DC 12 V ... 24 V |
| Warm-up time: | ? 5 min |
| Indication: | LEDs, 4'' color-display on optional controller unit |
| Weight 3): | 250 g |
| Housing material: | Aluminum housing with glass lens |
| footnote | 1) Extenable by cable to max. 50 m 2) DC 12 V (-5 %) ... DC 24 V (+10 %) 3) Inclusive 0.5 m cable |
| Enclosure rating: | IP 67 |
|---|---|
| Protection class: | III |
| Ambient temperature: | Operation: -10 ... +50 °C, Storage: -20 ... +60 °C |
| Temperature drift: | ± 0.01 % FS/K (FS = Full Scale = Measuring range of sensor) |
| Vibration resistance: | 10 Hz ... 55 Hz (amplitude 1.5 mm, x-, y-, z-axis 2 hours each) |
| Shock resistance: | 50 G (x-, y-, z-axis 3 times each) |
| Relative humidity (not condensing): | 35 % ... 85 % |
| Typ. ambient light safety: | Artificial light: ? 3,000 lx, Sunlight: ? 10,000 lx |